Accuracy validation of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure

Research output: Contribution to journalArticlepeer-review

29 Citations (Scopus)

Abstract

Objective To assess the accuracy of the Microlife 3AS1-2 blood pressure (BP) device in pregnant women with low BP to investigate suitability for hypotensive detection in low-income and middle-income countries.MethodsA prospective observational study was carried out evaluating the Microlife 3AS1-2, a hand-held, upper-arm, semiautomated BP device, according to British Hypertension Society (BHS) protocol methods. Thirty (stable) pregnant women with a clinical systolic BP less than 100 mmHg and/or diastolic BP less than 60 mmHg were recruited from antenatal wards and clinics and their BP was measured by three trained observers at a district-level hospital in South Africa. Accuracy was assessed according to the BHS grading criteria (A/B=pass) and the ANSI/AAMI/ISO standard for mean difference and SD (≤5±8 mmHg).ResultsThe device achieved an A/A grade according to the BHS grading criteria. The mean difference±SD between the observer and the test device was 0.5±6.2 and 1.3±5.4 mmHg for systolic and diastolic BP, respectively, fulfilling the standard required by the ANSI/AAMI/ISO protocol. All observer differences were within 4 mmHg.ConclusionAccording to the BHS protocol, the Microlife 3AS1-2 BP device is accurate in pregnant women with low BP. The device has been validated previously in pregnancy and pre-eclampsia and also fulfils the criteria of the WHO for use in a low-resource setting. Although unstable women were not included in this validation (for safety and pragmatic reasons), this device could potentially improve the detection of shock secondary to obstetric haemorrhage or sepsis, as well as being used in pre-eclampsia, particularly in low-income and middle-income countries.

Original languageEnglish
Pages (from-to)299-302
Number of pages4
JournalBLOOD PRESSURE MONITORING
Volume20
Issue number5
DOIs
Publication statusPublished - 23 Sept 2015

Keywords

  • blood pressure measurement
  • hypotension
  • low-resource setting
  • pregnancy
  • validation

Fingerprint

Dive into the research topics of 'Accuracy validation of the Microlife 3AS1-2 blood pressure device in a pregnant population with low blood pressure'. Together they form a unique fingerprint.

Cite this