Coil-based artifact reduction

Research output: Contribution to journalArticlepeer-review

27 Citations (Scopus)

Abstract

Multiple MRI receiver coils provide extra information and can enable the reconstruction of multiple images using data from different combinations of coils. Comparison of these images shows that artifacts due to motion or flowing blood appear with different intensities due to the differing coil sensitivities. Typically, the artifact appears amplified in regions of low coil sensitivity. An optimization routine was developed to correct for the artifact by comparing reconstructions from various coil combinations and favoring a self-consistent solution. It is demonstrated that images artifacted by blood flowing in the aorta, or translational motion of the head, can be improved. (C) 2004 Wiley-Liss, Inc.
Original languageEnglish
Pages (from-to)825 - 830
Number of pages6
JournalMagnetic Resonance in Medicine
Volume52
Issue number4
DOIs
Publication statusPublished - Oct 2004