Abstract
This study compared a laser fluorescence (LF) system (DIAGNOdent) with a visual caries scoring system for in vivo detection and diagnosis of occlusal caries under the conditions of an epidemiological study, in 132 mandibular and 38 maxillary first permanent molars in 170 children (mean age: 6.85 +/- 0.58 years). The teeth were cleaned and occlusal caries status in a selected investigation site recorded using both visual and LF systems. The LF readings were interpreted both according to the cut-off points recommended by the manufacturers and those based on laboratory research with histological validation. The percentage agreement of the LF and visual scoring system was better using the cut-off limits recommended by the manufacturer. Histological validation was not possible in this clinical study, but it appeared that either the LF method was overscoring some lesions or the visual method was underscoring them. Since the LF instrument cannot be expected to differentiate caries from hypomineralisation, it should probably be used as an adjunct to a clinical examination. Copyright (C) 2001 S. Karger AG, Basel.
Original language | English |
---|---|
Pages (from-to) | 421 - 426 |
Number of pages | 6 |
Journal | Caries research |
Volume | 35 |
Issue number | 6 |
DOIs | |
Publication status | Published - Nov 2001 |