Dentinal temperature transients caused by exposure to CO2 laser irradiation and possible pulpal damage

I. W. M. Jeffrey, B. Lawrenson, E. M. Saunders, C. Longbottom

Research output: Contribution to journalArticlepeer-review

37 Citations (Scopus)

Abstract

An investigation is described that attempts to establish, in vitro, the characteristics of heat transference following laser irradiation of bovine dentinal tissue and the relationship with the periodicity of radiation. The results of this study appear to indicate that at depths of overlying dentine of up to 3 mm, laser-induced thermal injury to the pulp is a definite possibility. Fail-safe facilities to prevent build up of heat must be incorporated into the design of dental lasers to allow their beneficial effects to be utilized without the risk of iatrogenic damage.

Original languageEnglish
Pages (from-to)31-36
Number of pages6
JournalJournal of Dentistry
Volume18
Issue number1
DOIs
Publication statusPublished - Feb 1990

Keywords

  • Animals
  • Cattle
  • Dental Pulp
  • Dentin
  • Dose-Response Relationship, Radiation
  • Hot Temperature
  • Lasers

Fingerprint

Dive into the research topics of 'Dentinal temperature transients caused by exposure to CO2 laser irradiation and possible pulpal damage'. Together they form a unique fingerprint.

Cite this