Does the Instantaneous Wave-Free Ratio Approximate the Fractional Flow Reserve?

Nils P. Johnson, Richard L. Kirkeeide, Kaleab N. Asrress, William F. Fearon, Timothy Lockie, Koen M. J. Marques, Stylianos A. Pyxaras, M. Cristina Rolandi, Marcel van 't Veer, Bernard De Bruyne, Jan J. Piek, Nico H. J. Pijls, Simon Redwood, Maria Siebes, Jos A. E. Spaan, K. Lance Gould*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

95 Citations (Scopus)

Abstract

Objectives This study sought to examine the clinical performance of and theoretical basis for the instantaneous wave-free ratio (iFR) approximation to the fractional flow reserve (FFR).

Background Recent work has proposed iFR as a vasodilation-free alternative to FFR for making mechanical revascularization decisions. Its fundamental basis is the assumption that diastolic resting myocardial resistance equals mean hyperemic resistance.

Methods Pressure-only and combined pressure-flow clinical data from several centers were studied both empirically and by using pressure-flow physiology. A Monte Carlo simulation was performed by repeatedly selecting random parameters as if drawing from a cohort of hypothetical patients, using the reported ranges of these physiologic variables.

Results We aggregated observations of 1,129 patients, including 120 with combined pressure-flow data. Separately, we performed 1,000 Monte Carlo simulations. Clinical data showed that iFR was +0.09 higher than FFR on average, with +/- 0.17 limits of agreement. Diastolic resting resistance was 2.5 +/- 1.0 times higher than mean hyperemic resistance in patients. Without invoking wave mechanics, classic pressure-flow physiology explained clinical observations well, with a coefficient of determination of >0.9. Nearly identical scatter of iFR versus FFR was seen between simulation and patient observations, thereby supporting our model.

Conclusions iFR provides both a biased estimate of FFR, on average, and an uncertain estimate of FFR in individual cases. Diastolic resting myocardial resistance does not equal mean hyperemic resistance, thereby contravening the most basic condition on which iFR depends. Fundamental relationships of coronary pressure and flow explain the iFR approximation without invoking wave mechanics. (J Am Coll Cardiol 2013;61:1428-35) (C) 2013 by the American College of Cardiology Foundation

Original languageEnglish
Pages (from-to)1428-1435
Number of pages8
JournalJournal of the American College of Cardiology
Volume61
Issue number13
DOIs
Publication statusPublished - 2 Apr 2013

Keywords

  • coronary physiology
  • fractional flow reserve
  • instantaneous wave-free ratio
  • Monte Carlo simulation
  • myocardial resistance
  • vasodilation
  • STENOSIS SEVERITY
  • CORONARY
  • INTRACORONARY
  • ADENOSINE

Fingerprint

Dive into the research topics of 'Does the Instantaneous Wave-Free Ratio Approximate the Fractional Flow Reserve?'. Together they form a unique fingerprint.

Cite this