Fluoroscopic Guide to an Ideal Implant Position for Sapien XT and CoreValve During a Valve-in-Valve Procedure

Vinnie N. Bapat*, Rizwan Q. Attia, Fortunata Condemi, Ravi Visagan, Maya Guthrie, Shelina Sunni, Martyn Thomas

*Corresponding author for this work

    Research output: Contribution to journalArticlepeer-review

    38 Citations (Scopus)

    Abstract

    Objectives 
    This study sought to provide a guide to the fluoroscopic appearances of various valve-in-valve (VIV) combinations by deploying a transcatheter heart valve (THV) within a degenerated surgical heart valve (SHV) in an ideal position.

    Background 
    VIV procedures are being increasingly performed with substantial experience acquired in treating degenerated SHV in the aortic position with Sapien/Sapien XT (Edwards Lifesciences Ltd., Irvine, California) and CoreValve/Evolute (Medtronic Inc., Minneapolis, Minnesota) valves. Although less invasive than conventional surgery, securing the THV in an optimal position within the SHV determines the success of this novel treatment.

    Methods 
    For VIV implantation, we selected appropriate Sapien XT and CoreValve/Evolute sizes depending on the internal diameter of the SHV. Implantation was performed in vitro. In case of the Sapien XT valve, it was deployed 4 to 5 mm below the sewing ring of the SHV, whereas the CoreValve/Evolute was deployed 5 mm below the level of the sewing ring. Photographs and fluoroscopic images of the various VIV combinations were obtained in side profile to study the ideal position and end-on profile to study the circularity of the THV.

    Results 
    Fluoroscopic images obtained in side profile highlighted the differences in various VIV combinations, as all SHV are unique in their fluoroscopic appearances. Also, all THV implants in various VIV combinations achieved a nearly circular shape.

    Conclusions 
    To achieve an optimal result when considering VIV, it is important to be familiar with the structure and fluoroscopic appearances of the failed SHV, the THV used, and their combination.

    Original languageEnglish
    Pages (from-to)1186-1194
    Number of pages9
    JournalJacc-Cardiovascular Interventions
    Volume6
    Issue number11
    DOIs
    Publication statusPublished - Nov 2013

    Keywords

    • bioprosthesis
    • transcatheter aortic valve implantation
    • transcatheter valve
    • valve-in-valve
    • valve replacement
    • TRANSCATHETER
    • BIOPROSTHESES
    • REPLACEMENT