Abstract
Aims: We sought to evaluate the first-in-man use of a new system for implantation of covered stents in patients with complex structural and CHD. Methods and results: Retrospective data were collected of the first 13 NuDEL™ delivery systems used in patients. The NuDEL™ comprises a covered Cheatham-Platinum stent mounted on a balloon-in-balloon and pre-loaded in a long delivery sheath. Data were collected from three centres in the United Kingdom and Ireland. A total of 13 covered stents were delivered via 12 NuDEL™ delivery systems in 12 patients. Among them, six patients had coarctation of the aorta, five patients had right ventricular outflow tract stenosis, and one patient had severe stenosis of a Mustard systemic venous baffle. There were no complications, and all the stents were deployed in the desired position with satisfactory haemodynamic results. Conclusions: The development of a bespoke system of a pre-mounted, pre-loaded covered stent may negate some of the technical challenges that complicate large-calibre stent deployment. Our preliminary results suggest that the NuDEL™ system is a safe and effective means of covered stent deployment in challenging anatomy.
Original language | English |
---|---|
Pages (from-to) | 1-5 |
Number of pages | 5 |
Journal | CARDIOLOGY IN THE YOUNG |
DOIs | |
Publication status | Accepted/In press - 15 Jun 2017 |
Keywords
- coarctation
- Covered stent
- percutaneous pulmonary valve
- right ventricular outflow tract stent