Abstract
We have carried out a PET performance evaluation of an SiPM based scanner designed for fully simultaneous preclinical PET/MR studies. The PET scanner has an inner diameter of 20 cm with a crystal size of 1.3 by 1.3 by 10 mm. The crystals are read out using MR-compatible SiPMs to allow the PET scanner to be located within a Philips 3T Achieva MRI scanner.
The spatial resolution of the system, measured using SSRB and 2D FBP is just under 2.4mm in the trans-axial and axial directions. The system sensitivity is 0.6 % for a point source at the centre of the field of view. The true coincidence count rate shows no sign of saturating at 30 MBq, at which point the randoms fraction is 9%, and the scatter fraction for a rat sized object is approximately 23%. Artefact-free images of phantoms have been obtained using SSRB/FBP and iterative reconstructions. The current performance is limited because only one of three axial ring positions is currently populated with detectors, and limitations of the first-generation detector readout ASIC used in the system.
The performance of the system as described is sufficient for imaging rat-sized animals and large organs within the mouse. We have demonstrated here the feasibility of using the system to investigate dynamic processes simultaneously in a mouse using PET and MR with a dual labeled PET/MR probe. Extrapolating from the current performance results we anticipate that population of all three detector rings and upgrading of the readout ASIC will result in an MR-compatible PET scanner with performance similar to that of state-of-the-art non-MR-compatible pre-clinical systems.
Original language | English |
---|---|
Title of host publication | 2012 IEEE Nuclear Science Symposium And Medical Imaging Conference (NSS/MIC) |
Editors | B Yu |
Place of Publication | Piscataway, N.J. |
Publisher | IEEE |
Pages | 2776-2779 |
Number of pages | 4 |
Volume | N/A |
Edition | N/A |
ISBN (Print) | 9781467320306 |
DOIs | |
Publication status | Published - 2012 |
Event | IEEE Nuclear Science Symposium / Medical Imaging Conference Record (NSS/MIC) / 19th Room-Temperature Semiconductor X-ray and Gamma-ray Detector Workshop - Anaheim, Canada Duration: 29 Oct 2012 → 3 Nov 2012 |
Publication series
Name | IEEE Nuclear Science Symposium Conference Record |
---|---|
Publisher | IEEE |
ISSN (Print) | 1082-3654 |
Conference
Conference | IEEE Nuclear Science Symposium / Medical Imaging Conference Record (NSS/MIC) / 19th Room-Temperature Semiconductor X-ray and Gamma-ray Detector Workshop |
---|---|
Country/Territory | Canada |
Period | 29/10/2012 → 3/11/2012 |