Scanning transmission x-ray microscopy with a configurable detector

A Gianoncelli, G R Morrison, B Kaulich, D Bacescu, J Kovac

Research output: Contribution to journalArticlepeer-review

105 Citations (Scopus)

Abstract

A configurable charge-coupled-device (CCD) detector is used in a scanning transmission x-ray microscope to record the transmitted intensity distribution for every pixel in a raster scan of the sample. Real-time processing of the CCD frames gives simultaneous absorption and phase contrast image signals from a single scan. The CCD combines fast frame-transfer readout with very high sensitivity, and makes use of x-ray to visible-light coupling to allow operation over a wide range of photon energies, from the oxygen K edge upwards. Tests on the Twinmic end-station at the Elettra synchrotron are reported.
Original languageEnglish
Article number251117
JournalAPPLIED PHYSICS LETTERS
Volume89
Issue number25
DOIs
Publication statusPublished - 2006