Simultaneous PET and MR imaging

Y. P. Shao, S. R. Cherry, K. Farahani, K. Meadors, S. Siegel, R. W. Silverman, P. K. Marsden

Research output: Contribution to journalArticlepeer-review

341 Citations (Scopus)

Abstract

We have developed a prototype PET detector which is compatible with a clinical MRI system to provide simultaneous PET and MR imaging. This single-slice PET system consists of 48 2 x 2 x 10 mm(3) LSO crystals in a 38 mm diameter ring configuration that can be placed inside the receiver coil of the MRI system, coupled to three multi-channel photomultipliers housed outside the main magnetic field via 4 m long and 2 mm diameter optical fibres. The PET system exhibits 2 mm spatial resolution, 41% energy resolution at 511 keV and 20 ns timing resolution. Simultaneous PET and MR phantom images were successfully acquired.
Original languageUndefined/Unknown
Pages (from-to)1965-1970
Number of pages6
JournalPhysics in Medicine and Biology
Volume42
Issue number10
DOIs
Publication statusPublished - 1997

Cite this