Tissue and device visualization using parallel transmission for interventional MRI

Research output: Chapter in Book/Report/Conference proceedingMeeting abstractpeer-review

Abstract

It is demonstrated that parallel transmit (PTx) can be used to both safely visualise tissue (no heating even from high SAR
sequences), and safely visualise a wire by direct coupling using low power MR sequences. A low power version of a TSE sequence was found to give the best depiction of the wire when using this method.
Original languageEnglish
Title of host publicationISMRM
Place of PublicationHonolulu HI
Publication statusPublished - 2017