Comparing approaches to correct for respiratory motion in NH3 PET-CT cardiac perfusion imaging

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)

Abstract

Aim: Respiratory motion affects cardiac PET-computed tomography (CT) imaging by reducing attenuation correction (AC) accuracy and by introducing blur. The aim of this study was to compare three approaches for reducing motion-induced AC errors and evaluate the inclusion of respiratory motion correction.
Materials and methods: AC with a helical CT was compared with averaged cine and gated cine CT, as well as with a pseudo-gated CT, which was produced by applying PET-derived motion fields to the helical CT. Data-driven gating was used to produce respiratory-gated PET and CT images, and 60 NH3 PET scans were attenuation corrected with each of the CTs. Respiratory motion correction was applied to the gated and pseudo-gated attenuation-corrected PET images.
Results: Anterior and lateral wall intensity measured in attenuation-corrected PET images generally increased when PET-CT alignment improved and decreased when alignment degraded. On average, all methods improved PET-CT liver and cardiac alignment, and increased anterior wall intensity by more than 10% in 36, 33 and 25 cases for the averaged, gated and pseudo-gated CTAC PET images, respectively. However, cases were found where alignment worsened and severe artefacts resulted. This occurred in more cases and to a greater extent for the averaged and gated CT, where the anterior wall intensity reduced by more than 10% in 21 and 24 cases, respectively, compared with six cases for the pseudo-gated CT. Application of respiratory motion correction increased the average anterior and inferior wall intensity, but only 13% of cases increased by more than 10%.
Conclusion: All methods improved average respiratory-induced AC errors; however, some severe artefacts were produced. The pseudo-gated CT was found to be the most robust method. 

Original languageEnglish
Article numberN/A
Pages (from-to)1174-1184
Number of pages11
JournalNuclear Medicine Communications
Volume34
Issue number12
DOIs
Publication statusPublished - Dec 2013

Keywords

  • PET-computed tomography
  • respiratory motion
  • respiratory motion correction
  • ATTENUATION-CORRECTION
  • GATED PET
  • INTERNAL ANATOMY
  • RECONSTRUCTION
  • IMAGES
  • ARTIFACTS
  • LUNG
  • TOMOGRAPHY
  • MODEL
  • REGISTRATION

Fingerprint

Dive into the research topics of 'Comparing approaches to correct for respiratory motion in NH3 PET-CT cardiac perfusion imaging'. Together they form a unique fingerprint.

Cite this